Introduction

Keratometry (K) has been traditionally calculated by means of the keratometric index, which enables us to estimate the dioptric power of the whole cornea from measurements of the anterior corneal surface only [1]. More recently, Scheimpflug cameras have allowed us to measure the curvature of the posterior corneal surface as well, resulting in a more accurate calculation (rather than just an estimation) of total corneal power (TCP). Several studies, however, have demonstrated that TCP by Scheimpflug cameras is lower than the value given by standard K [2,3,4,5,6,7,8]. This discrepancy has limited its adoption for intraocular lens (IOL) power calculation, since specifically calculated optimized IOL constants would be required in order to adjust for differences [2].

In 2019 a new approach was offered by the IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), which can calculate Total Keratometry (TK) from both corneal surfaces and adjust this value so that, on average, it does not differ from standard K in normal unoperated eyes. TK has been investigated by a few authors for IOL power calculation accuracy in patients undergoing cataract surgery with or without toric IOLs and in eyes with previous corneal refractive surgery [9,10,11,12]. However, two issues have not yet been assessed and this paper aims to clarify them. Our goals were as follows: (1) to measure the repeatability of TK as compared to standard K measurements in previously unoperated eyes undergoing cataract surgery and in eyes with previous myopic corneal laser surgery and (2) to compare corneal astigmatism measured by TK to the total corneal astigmatism (TCA) measured by a Scheimpflug camera in patients scheduled for cataract surgery.

Methods

Patients

This was a prospective case series. The study protocol was approved by the Ethics Committee of G.B. Bietti Foundation IRCCS, Rome, Italy and complied with the tenets of Declaration of Helsinki. All patients gave their informed consent.

Two groups were enrolled:

  • Group A included a consecutive series of unoperated eyes undergoing cataract surgery;

  • Group B included a consecutive series of eyes that had undergone either myopic PRK or femtoLASIK and were visited for a routine postoperative examination;

Both Groups were used to assess the repeatability of K and TK. Moreover, in Group A we investigated agreement between TK astigmatism and TCA measured by a Scheimpflug camera (Pentacam AXL, Oculus, Wetzlar, Germany). Exclusion criteria were as follows: a history of corneal disease or trauma, contact lens use in the past month, a previous diagnosis of keratoconus, suspect keratoconus or dry eye and any kind of ocular surgery (other than excimer laser in Group B).

Instruments

The IOLMaster 700 (software version 1.80), whose technology is based on swept-source optical coherence tomography (SS-OCT), was used to measure K and TK. With each scan, the device automatically acquires 3 B-scans of the cornea in 6 meridians, so that 18 measurements are taken altogether. Standard K is derived from 15 telecentric measurements of three rings, whose diameter varies slightly depending on the corneal curvature (for an average cornea with a radius of 7.7 mm, they are ~1.5, 2.5, and 3.3 mm). The average values obtained with this method are systematically identical to those taken at 2.5 mm with the IOLMaster 500. In the event of poor consistency among the three rings, with deviations from the inner, middle, and outer zones, a warning is shown (or the measurement is not displayed). The 1.3375 keratometric index was selected. As regards the posterior corneal curvature, the IOLMaster 700 first builds a toric anterior surface model from the telecentric 3-zone K and then measures pachymetry using SS-OCT in the six meridians. The pachymetry values are fitted to the anterior surface model to create the toric posterior surface model. The TK is calculated from the anterior and posterior corneal curvatures, as well as corneal thickness, by means of a thick lens formula. The TK values are adjusted with a proprietary method to match, on average, the K values, so that existing IOL calculation formulas and IOL constants can be adopted with no changes. For both K and TK, we recorded the powers of the flat and steep meridians, their average and their difference, i.e., the measured corneal astigmatism.

The Pentacam AXL (software version 1.21r43), which was used to measure the TCA, is a rotating Scheimpflug camera combined with partial coherence interferometry [13]. TCP and astigmatism are calculated by ray tracing through the anterior and posterior corneal surfaces [4]. In order to facilitate the comparison between the two devices, we selected the measurements centered on the corneal vertex and investigated those with a diameter of 2.0 and 3.0 mm. We analysed both the “ring” and “zone” measurements: the former provide curvature data along the 2.0 and 3.0 mm rings, whereas the latter also include information about the area within the measured ring.

Measurement procedures

Each device was brought into focus, and the eye was aligned along the visual axis by means of a central fixation light. Patients were asked to blink before each measurement. Both instruments were calibrated at the beginning of each day.

In order to assess the repeatability of K and TK measurements by the IOLMaster 700, three repeated consecutive measurements were acquired by one experienced examiner in both groups. The patients were asked to sit back after each measurement, and the device was realigned before the subsequent measurement. We checked that no alerts (!) had been displayed by the instrument software in order to assure the good quality of the measurements.

With the rotating Scheimpflug camera, only one measurement was taken, on condition that the quality specification was acceptable. In Group A, the order sequence of the devices was established at random.

Astigmatism analysis

Vector analysis was performed according to Næser [14], who described two equations for converting net corneal astigmatism in conventional script notation (M at α) into polar values:

$${\mathrm{KP}}\left( 0 \right) = {M} \times {\mathrm{cos}}\left( {2 \times {\upalpha}} \right),$$
(1)
$${\mathrm{KP}}\left( {45} \right) = {M} \times {\mathrm{sin}}\left( {2 \times {\upalpha}} \right),$$
(2)

where KP(0) and KP(45) are, respectively, the polar value along the zero and the 45 degree meridians, M is the astigmatism magnitude and α is the orientation of the steep meridian. Polar values can be used to calculate the average or the difference of any corneal astigmatism; the resulting values can be reconverted to the conventional notation using the following equations:

$${\mathrm{Magnitude}} = \sqrt {{\mathrm{KP}}\left( 0 \right)^2 {\,}+{\,} {\mathrm{KP}}\left( {45} \right)^2}.$$
(3)
$${\mathrm{Angle}} = {\mathrm{arctan}}\left( {\frac{{{M} - {\mathrm{KP}}\left( 0 \right)}}{{{\mathrm{KP}}\left( {45} \right)}}} \right).$$
(4)

Using Eqs. 14 it was possible to compare the mean TCA measured by the Scheimpflug camera to the mean corneal astigmatism by TK (first measurement only). For this purpose, only eyes with KA > 0.5 D in all three measurements by the IOLMaster 700 were analyzed (as previously done by Ventura et al.) [15], since eyes with a lower astigmatism magnitude are unlikely to undergo toric IOL implantation and are known to display larger variability in astigmatism axis measurements [16, 17]. Moreover, in order to better investigate the difference in corneal astigmatism between K and TK, a third group (Group A1) of unoperated eyes with KA > 0.5 D was enrolled (with the same exclusion criteria of Group A). This group underwent one measurement only and was not used to test the repeatability.

Vector analysis was carried out separately in eyes with with-the-rule astigmatism (WTR, steep corneal meridian between 60° and 120°), against-the-rule astigmatism (ATR, steep corneal meridian between 0° and 30° and between 150° and 180°) and oblique astigmatism (for the remaining orientations of the steep meridian).

Statistical analysis

Statistics were analyzed by means of MedCalc (version 12.3.0.0, MedCalc Software Ltd, Ostend, Belgium).

The analysis of repeatability was conducted according to the standards of the International Organization for Standardization [18], as recently reported for a similar study by our group [19]. Repeatability was assessed on the basis of intrasession test–retest variability, the coefficient of variation (COV), and the intraclass correlation coefficient (ICC).

The following methods were used:

  1. (1)

    Intrasession test–retest variability (also known as repeatability or limits of repeatability). This was calculated by multiplying the pooled within-subject SD (Sw) by 2.77 [20]. Based on repeatability, we can expect that the difference between two measurements for the same subject will be less than 2.77 Sw for 95% of paired observations.

  2. (2)

    COV. This was calculated as Sw divided by the mean of the measurements and expressed as a percentage. The COV was not calculated for parameters with both positive values and negative values (e.g., KP(0) and KP(45)) [21].

  3. (3)

    ICC. This is the ratio of the between-subjects variance to the sum of the pooled within-subject variance and the between-subjects variance. The ICC, which approaches 1.0 when there is no variance between repeated measurements, was automatically calculated using the two-way mixed model and absolute agreement. ICCs ranging from 0 to 1 are classified as follows: ICC less than 0.75 = poor agreement; ICC 0.75 to less than 0.90  = moderate agreement and ICC 0.90 and more = high agreement [22].

The variances of K and TK values were compared in both Groups with the F-test for variances. Linear regression was used to analyze the relationship between the difference K–TK and the value of K as well as between astigmatism magnitude and COV.

Based on the article by McAlinden [23], a minimum sample size of 43 eyes was estimated for the two groups used to assess repeatability: this sample size allowed us to have a 0.15 confidence in the estimate with three repeated measures. Actually, since we enrolled slightly larger samples, a 0.12 confidence in the estimate was calculated for Group A and 0.14 for Group B. These values are similar to or higher than those previously reported by other authors investigating the repeatability of corneal topographers, tomographers and keratometers [24,25,26].

Results

The analysis of repeatability was performed on 69 unoperated eyes of 69 consecutive patients (mean age: 68.8 ± 11.9 years; 38 females) and 51 post-excimer laser surgery eyes of 51 consecutive patients (mean age: 40.9 ± 11.4 years; 31 males). In Group A the mean K and the mean TK (first measurement only for both values) were, respectively, 43.14 ± 1.37 D and 43.18 ± 1.37 D; the mean of the differences between K and TK was statistically (p = 0.0006) but not clinically significant. The difference between the two corneal power calculation methods (K minus TK) ranged from −0.27 to +0.26 D and was not correlated to the magnitude of K (p > 0.05). In Group B, the mean K and the mean TK were, respectively, 40.14 ± 2.20 D and 39.71 ± 2.35 D; the mean of the differences between K and TK was statistically (p < 0.0001) and clinically significant, as it revealed that K overestimates corneal power compared to TK. The difference between the two corneal power calculation methods (K − TK) ranged from −0.06 to +0.88 D and was correlated to the magnitude of K (r = −0.6230; r2 = 0.3881; p < 0.0001), thus suggesting that flatter corneas had a higher difference between K and TK.

Repeatability analysis

Table 1 shows the test–retest variability, COV and ICC for K and TK. All values revealed high repeatability with ICC > 0.9; astigmatism measurements were slightly less repeatable, even if we limited their analysis to cases in which the K astigmatism was >0.50 D in all three measurements. Table 1 shows two interesting results. First, the repeatability of TK values is very close to that of K values in both unoperated corneas and post-excimer laser surgery corneas. Second, the repeatability of K and TK measurements does not get worse, but slightly improves in eyes with previous excimer laser surgery.

Table 1 Repeatability analysis of K and TK measurements provided by the IOLMaster 700.

Table 2 shows the results of the variance ratio F-test performed to compare the repeatability of K and TK within each sample. It may be observed that for most measurements the variance of K and TK did not show any statistically significant difference, the only exception being the measurements of the steep corneal meridian and the axis of TK astigmatism in post-excimer laser surgery eyes.

Table 2 Variance ratio (p value), as calculated by means of the F-test, between K and TK.

Table 3 shows the results of the F-test performed to compare the repeatability of K and TK between unoperated eyes and post-excimer laser surgery eyes. It shows the lack of statistically significant differences for most measurements; a few exceptions were observed for TK, where the variance of the steep and the average corneal power and the variance of the astigmatism axis was higher in post-excimer laser group (although these differences are clinically negligible, as Table 1 shows).

Table 3 Variance ratio (p value) as calculated by means of the F-test.

Astigmatism analysis

In Group A, the repeatability of the measurement of the K astigmatism magnitude tended to improve as the magnitude of astigmatism increased (r = −0.5065, r2 = 0.2565, p < 0.0001, Fig. 1a); this holds true for TK astigmatism in unoperated eyes (r = −0.5279, r2 = 0.2786, p < 0.0001, Fig. 1b), K astigmatism in post-refractive surgery eyes (r = −0.4895, r2 = 0.2396, p = 0.0002, Fig. 1c) and TK astigmatism in post-refractive surgery eyes (r = −0.5563, r2 = 0.3094, p < 0.0001, Fig. 1d).

Fig. 1: Linear regression between the coefficient of variation (COV) and the magnitude of astigmatism.
figure 1

A K astigmatism in unoperated eyes. B TK astigmatism in unoperated eyes. C K astigmatism in post-refractive surgery eyes. D TK astigmatism in post-refractive surgery eyes.

Due to the lower repeatability in eyes with a lower astigmatism magnitude, as previously stated, we performed a more detailed analysis in unoperated eyes whose K astigmatism magnitude was >0.5 D. This analysis was carried out in eyes of Group A and Group A1 (the latter included 83 eyes of 83 patients with a mean age of 60.6 ± 19.9 years) and involved 40 eyes with ATR astigmatism, 59 eyes with WTR astigmatism and 22 eyes with oblique astigmatism. As expected, in ATR eyes TK astigmatism (1.37 D @ 176.7°) was higher than K astigmatism (1.08 D @ 175.3°), whereas in WTR eyes TK astigmatism (1.16 D @ 91.0°) was lower than K astigmatism (1.35 D @ 91.8°); a small difference was observed also in eyes with oblique astigmatism (TK = 0.08 D @66.0°, K = 0.25 D @89.5°). Accordingly, the analysis of the first measurement showed that the vector KP(0) was higher with TK than with K in ATR eyes (1.36 ± 0.64 vs. 1.09 ± 0.64, p < 0.0001), in WTR eyes (−1.16 ± 0.92 vs −1.34 ± 0.89, p < 0.0001), and in eyes with oblique astigmatism (−0.00 ± 0.60 vs −0.20 ± 0.59, p < 0.0001). No statistically significant differences were obtained for KP(45) in any subgroup, meaning that there were no axis differences between K and TK astigmatism.

The comparison between TK astigmatism and TCA as measured by the Scheimpflug camera (which was possible in 35 out of 38 eyes due to a low quality of Scheimpflug measurements in 3 eyes) revealed that the measurements provided by the two technologies are not interchangeable (Table 4). In ATR eyes (n = 14), the mean TK astigmatism was higher than any TCA measurement. In WTR eyes (n = 10), the mean TK astigmatism was lower than any TCA measurement. In eyes with oblique astigmatism (n = 11), the mean TK astigmatism was slightly lower than all TCA measurements, but TCA at 3 mm apex/ring. Overall, the 3 mm apex/zone by the Scheimpflug camera gave the lowest difference between TK astigmatism and TCA in eyes with ATR astigmatism (0.09 D @ 10.9°) and oblique astigmatism (0.06 D @ 34.5°); the 2 mm apex/zone showed the lowest difference between TK and TCA in eyes with WTR astigmatism (0.19 D @ 35.6°).

Table 4 Mean net astigmatism values of total corneal astigmatism measured by TK and Scheimpflug imaging.

Table 5 shows the mean vectorial values of astigmatism measurements given by the two devices. The only statistically significant differences were observed for KP(45) vectors in ATR and WTR eyes, meaning a difference in the orientation of astigmatism.

Table 5 Mean vectorial values of total corneal astigmatism measured by TK and Scheimpflug imaging.

Discussion

Our data show a high repeatability of both K and TK values, as measured by the IOLMaster 700. The most important finding is that the repeatability of TK is as high as the repeatability of K in both unoperated and post-LASIK eyes. The relevance of TK is mainly related to the IOL power calculation in post-refractive surgery and in astigmatic eyes [9, 10, 12]. By taking the posterior corneal curvature into account, TK (unlike K) does not make any assumption about the anterior to posterior curvature ratio, which leads to the so-called “index of refraction error” in post-LASIK eyes [27], and reduces the overestimation of WTR astigmatism and the underestimation of ATR astigmatism.

The high repeatability of TK in unoperated eyes reported in this study confirms previous finding on 93 healthy volunteers (mean age = 52 years; range = 18–85 years), even if we observed a slightly higher Sw (probably related to the older age of the our sample, which may lead to a lower degree of cooperation) [28]. As regard other instruments, the Sw of TK in our study (0.09 D) lies between the values previously obtained by Aramberri et al. for the TCP measured by a rotating Scheimpflug camera (Pentacam HR, Oculus) and a dual Scheimpflug analyzer (Galilei G2, Port, Ziemer); they reported values of 0.05 and 0.12 D, respectively, in a sample of young patients (mean age = 35 years) [24]. However, comparison between studies have to be taken with caution due to the different study populations; measurements in older patients with cataract usually show a worse repeatability than those in young healthy subjects [13]. The main advantage compared to the TCP measured by the Scheimpflug systems is that TK can be directly entered into IOL power formulas, since in unoperated eyes its mean value (43.18 ± 1.37 D) does not show any clinically significant difference compared to standard K (43.14 ± 1.37 D), as was previously found also by Srivannaboon et al [9]. Moreover, the maximum difference between K and TK in unoperated eyes is never higher than a quarter of a diopter. In contrast, the TCP measured by any Scheimpflug camera is lower compared to K and requires specifically optimized constants [4, 19, 26, 29].

We also aimed to assess whether the repeatability of K and TK decreased in post-LASIK eyes, as ocular surface changes may affect the quality of measurements (the repeatability of simulated K has been reported to be lower after excimer laser with both Scheimpflug and OCT systems) [19, 26]. However, we observed that the repeatability was slightly better: this finding may depend on the different population of unoperated and post-LASIK patients (rather than on the corneal conditions), as post-LASIK patients are younger and thus able to maintain better fixation. Interestingly, we observed that, in eyes with previous myopic surgery, the difference between K and TK was correlated to the magnitude of K: the lower the K, the larger the difference. This is due to the fact that, after myopic PRK or LASIK, corneal power is overestimated by K; the overestimation is greater for higher degrees of myopic correction.

As regards the repeatability of K in unoperated eyes, several authors had already reported that the values measured by the IOLMaster 700 show high repeatability, with Sw ranging between 0.07 and 0.13 D [30,31,32]. A similar repeatability was reported for two optical biometers with a Placido disc, the Aladdin (Topcon Europe, Florence, Italy) and the OA-2000 (Tomey, Nagoya, Japan), whose values of Sw for K were 0.09 and 0.06 D, respectively [33, 34]. The repeatability of K measurements (Sw = 0.09 D) by the Lenstar (Haag-Streit, Bern, Switzerland) was also close to that found in the present study [32]. If we specifically consider cataract patients, the repeatability of K (0.08 D) and TK (0.09 D) was better than the repeatability of K as measured by the Pentacam AXL (Sw = 0.13 D) and the Allegro Biograph (Sw = 0.16 D), which is a rebranded LenStar [13]. Of course, data from different studies should be compared to caution, since they are influenced by the different populations analysed and the different environmental conditions.

Astigmatism analysis showed a good repeatability in both groups (the ICC of magnitude, axis and both vectors was >0.9), although to a slightly lesser extent compared to steep, flat, and average corneal power measurements, as also found by other authors who reported lower repeatability for astigmatism compared to average corneal power [15, 17, 24]. The repeatability was worse for eyes with lower amounts of astigmatism, in agreement with the results reported by Kim et al [17]. In unoperated eyes, the Sw of the K (0.13 D) and TK (0.14 D) astigmatism magnitude was similar to or better than that previously reported for other instruments, such as the LenStar (Sw = 0.13), the Cassini (i-Optics, the Netherlands, Sw = 0.31) and the Galilei G4 (0.18 D) [15, 17].

Vectorial analysis in unoperated eyes confirmed that, compared to TK, K underestimates corneal astigmatism in eyes with ATR astigmatism and overestimates it in eyes with WTR astigmatism. This result, which has not been previously reported in relation to measurements by the IOLMaster 700, mirrors previous experiences with other technologies measuring posterior and TCA [35,36,37]. Moreover, a comparison with the TCA measurements by the Pentacam AXL revealed clinically significant differences (mainly due to the KP(45) vector, i.e., to the torsional effect of astigmatism). Hence, as previously observed, TCA measurements provided by the two devices cannot be considered interchangeable [28]. We can speculate that these discrepancies are related to the different methods used to calculate TCA, since previous studies did not find significant differences between the two devices in the measurement of keratometric astigmatism [38, 39]. An analysis of a larger sample is warranted in this regard, since the comparison was performed on only a few eyes.

This study has some limitations. First, we did not investigate repeatability in specific conditions, such as keratoconus, for which further investigations are necessary. Second, we did not assess the accuracy of K and TK for IOL power calculation in unoperated eyes: this issue, which has already been evaluated in a few relatively small studies [9, 11], will be the subject of future projects involving a larger sample. Third, the comparison of the TCA by the two devices was performed on a relatively low number of cases. However, our results agree with a previous larger study [25].

In conclusion, our data support the high repeatability of K and TK measurements by the IOLMaster 700 in both unoperated eyes and post-excimer laser surgery eyes. When compared to conventional K, TK confirmed to provide similar spherical equivalent measurements in unoperated eyes, lower spherical equivalent measurements in post-myopic laser surgery eyes and different astigmatism measurements. Moreover, TCA provided by total keratometry and Scheimpflug imaging are not interchangeable.

Summary

What was known before

  • Total Keratometry (from both corneal surfaces) and standard Keratometry (from anterior corneal surface only) by IOLMaster show on average similar values in unoperated eyes.

What this study adds

  • Total Keratometry measurements show high repeatability in both unoperated and post-LASIK eyes. Such repeatability is as high as the repeatability of standard keratometry.

  • Total corneal astigmatism, as measured by Total Keratometry, is different compared to total corneal astigmatism measured by a Scheimpflug camera.